NONMETALLIC TUBING ASSEMBLY. - Continued
FLOW ADJUST REGULATING VALVE.

M140 Chemical Agent Alarm \(Nsn 6665-01-083-2749\)
Page Navigation
  89    90    91    92    93  94  95    96    97    98    99  
img
. .
TM 3-6665-329-13&P
This task covers:
Removal.
a.
Installation.
b.
INITIAL SETUP
Tools:
Electronic equipment tool
kit TK-105/G
Equipment condition: Lid and panel assembly removed during troubleshooting.
Removal.
a.
(1) Unscrew two nuts (2) and remove nonmetallic tubing (1) from PRESSURE LOCK
angle valve (3).
(2) Place PRESSURE LOCK angle valve handle (4) in VERT position (straight up).
(1) Place replacement PRESSURE LOCK angle valve handle (4) in VERT
(straight up) position.
(2) Remove two nuts (2) and ferrules from replacement PRESSURE LOCK angle valve.
(3) Install PRESSURE LOCK angle valve (3), and secure with retaining nut (5).
(4) Install nonmetallic tubing (1) and tighten nuts (2).
3-73