TEST SET CASE - Continued
TEST SET CASE - Continued

M140 Chemical Agent Alarm \(Nsn 6665-01-083-2749\)
Page Navigation
  65    66    67    68    69  70  71    72    73    74    75  
img
.
TM 3-6665-329-13&P
c.
Installation.
(1)
Unfasten four fasteners (5) on new case.
(2)
Lift lid (6) and slide off of hinges (7).
Lay lid aside until needed.
(3)
Insert one hand into flowmeter opening and place fingers around right flowmeter bracket (13.5).
(4)
(5)
Grasp binding posts (14) with other hand.
BE SURE NONMETALLIC TUBING IS NOT BETWEEN PANEL ASSEMBLY AND CASE CENTER
SUPPORT.
NOTE
To place panel assembly into case, be sure flowmeter is on the hinge side of case.
Place panel assembly (15) into case (1).
(6)
Lower flowmeter (12) and finger tighten two screws (11).
(7)
Secure to case with 21 flatwashers (10), Iockwashers (9) and screws (8).
(8)
Push in and release three fasteners (16) and open accessory shelf (17).
(9)
(10)
Position filter holder (27) over mounting holes, secure with four Iockwashers (26) and screws (25).
(11)
Position clamp (24) in mounting position. Secure wtih one screw (21), washers (22), and locknut
(23).
(12)
Position clip (32) over mounting hole and between raised areas.
(13)
Secure clip with screw (31).
(14)
Repeat steps 12 and 13 for other slip.
(15)
Apply sealing, locking and retaining compound (item 2, app D) to threads of screws (28).
3-49