TEST SET CASE
TEST SET CASE - Continued

M140 Chemical Agent Alarm \(Nsn 6665-01-083-2749\)
Page Navigation
  63    64    65    66    67  68  69    70    71    72    73  
img
ARMY TM 3-6665-329-13&P
AIR FORCE TO 11H2-18-1
Removal.
b.
(1)
Unfasten four fasteners (5).
Lift lid (6) and slide off of hinges (7).
(2)
Lay lid aside until needed.
(3)
Remove 21 screws (8), Iockwashers (9), and flatwashers (10).
(4)
Unscrew two screws (11).
(5)
Raise flowmeter (12) and lock in position with hinge spring bracket (13).
(6)
Insert one hand into flowmeter opening and place fingers around right flowmeter
(7)
..
bracket (13.5).
(8) Grasp binding posts connectors (14) with other hand and lift panel assembly (15) out of case.
--------------
--C-A-U---I-O-N
-- -- --T -- --
BE CAREFUL NOT TO PINCH OR DAMAGE TUBING WHEN LAYING PANEL ASSEMBLY ASIDE.
3-47
Change 2