T.O. 33B-1-1Figure 1-20.Angle Beam Distance Calibration with IIW Block1.8.2.6.5AngleBeamDistanceCalibration(IIWBlock).The following procedure works only for a 45o angle beam.a.Position the, search unit at P-1 as shown in Figure 1-20 and adjust the position laterally to obtainequal amplitude signals from R-2 and R-4.b.The distance between the R-2 and R-4 signals is 2 inches. Set the time base range for the applicabledistance calibration.c.Another method is to position the search unit along scale ‘‘C’’ and obtain a peak signal from the holeas shown in Figure 1-20. Distance is read directly off the scale. P-2 shows 2.5 inches; P-3 shows 5inches.Change 31-67