T.O. 33B-1-1Figure 1-21.Angle Beam Distance Calibration with Miniature Angle Beam Block1.8.2.6.6AngleBeamDistanceCalibration(MiniatureAngleBeamBlock).The following procedure works only for a 45o angle beam.a.Position the search unit at P-1 and then P-2 as shown in Figure 1-21. Obtain peak signals from R-1and then from R-2.b.The angle beam metal travel at P-1 is 1.00 inch; at P-2, it is 2.00 inches. Use multiple back ref lectionsto calibrate longer distances.CAUTIONEnsure you are using the proper transducer and standard matched for thematerial you wish to inspect.EX: A = aluminum transducerS = steel transducer1-68Change 3