T.O. 33B-1-1are 0.25 and 0.5 inch thick, scale divisions at 2.5 and 5.0 could be used, or for better resolution 5.0 and10.0 could be used. In the first case, each major division on the scale equals 0.1 inch. In the secondcase, each major division equals 0.05 inch. In general, each major division equals 0.1 of the thicknessfor which the echo signal is placed at division 10.d.Obtain an echo signal from the thinner reference standard and adjust the instrument Delay control toposition the signal at the chosen scale division.e.Obtain an echo signal from the thicker reference standard and adjust the instrument Range control toposition the signal at the chosen respective scale division.f.Repeat paragraphs d and e until no further adjustment is necessary. Always place the same verticalposition of the signals at the respective divisional markings; that is, place the point where the signalbreaks the baseline, crosses the 10 percent vertical-scale line or any other convenient vertical-scaleline.g.Place the search unit on the test part and obtain a signal. Read the thickness from the horizontalscale of the display, using the same vertical location on the signal that was used when placing thesignals from the reference standards. On parts with irregular or pitted surfaces, take severalreadings and record the average. If the surface is curved, see paragraph 5.5.5.2 for additionalinstructions. As required, multiply the readings by a correction factor (see paragraph 5.5.3.3).h.The calibration should be rechecked each one-half hour during continuous use. If it is determinedthat the instrument is out of calibration, all measurements made since the last satisfactorycalibration should be repeated after correcting the calibration.SECTION VIULTRASONIC LEAK TESTING5.6ULTRASONICLEAKTESTING.5.6.1Introduction.NOTEStd cm3/s is a unit of gas leakage rate. One std cm3/s means that one cubiccentimeter of gas at atmospheric pressure (14.7 pounds per square inch) andstandard temperature (20oC) passes through the leak each second. A leakagerate may also be expressed in the derived SI units of pascal cubic meters persecond Pa m3/s). Adding the unit of pressure, pascal provides a valid leakagerate without having to convert to atmospheric pressure conditions. 1 Pa m3/s= 10 std cm3/s.a.Gas leakage of f low greater than about 0.1 standard cubic centimeters per second (std cm3/s) producesultrasonic energy with frequencies in the range of 30,000 to 50,000 Hz under most conditions. The gasleakage can be out of a pressure system or into a vacuum system. The ultrasonic leak detector utilizesa probe containing a microphone sensitive to this range of ultrasonic frequencies. The ultrasonicenergy from leakage is detected. Background noise at other frequencies does not affect the detector.b.Ultrasonic energy in this relatively low-frequency range of 30,000 to 50,000 Hz travels easily throughair. The leak detector can therefore detect leakage with the probe located away from the leak. Themaximum detection distance depends on the leakage rate. A leakage rate of 0.1 std cm3/s can be5-84Change 2
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business