TM 5-6635-386-12&PWhen pressed simultaneously with LOCK, will erase all four memories and willinitiate a 1 minute count into all four memories. The displays will each read 65536 ifthe test signal has been properly processed. This tests all digital circuitry in thetester, but does not test the analog circuitry (high voltage supplies).NOTEAfter approximately Jul 78, TEST will count 1/4 minute andwill display 16384. This shortens TEST time as aconvenience.D-ONLY=When pressed, the tester will count Density Only, retaining last moisture count. D-ONLY is cancelled by pressing an Operations Key or by pressing CLEAR.M-ONLY=Same as D-ONLY except that only moisture is counted.1/4 - MINInitiates 1/4 minute count cycle. Use for Statistical check per Paragraph(STAT-CHECK) =1-7.1/2 - MIN=Initiates 1/2 minute count cycle.1 - MIN=Initiates 1 minute count cycle.2 - MIN=Initiates 2 minute count cycle.PRECISIONOFVARIOUSTIMEPERIODS(Read Para 1-7 through 1-9)NOTE:ALLDISPLAYSNORMALIZEDTO1/4MINUTE.(in PCF1/41/212@ 120#)BS1.20.85.60.42AC1.00.71.50.356".50.35.25.18Moisture.50.35.25.18Legend:PCF - Pounds Per Cubic FootBS - BackscatterAC - Asphaltic Concrete1-5