T.O. 33B-1-1c.Move the reference standard toward the magnet until a half-scale meter reading is obtained. (If thesouth pole of the magnet is up, the meter needle will def lect to the right.)d.Mark this position on the sheet of paper by tracing along the bottom with a marker.e.Place an in-use field indicator in the exact position traced for the reference standard and note thereading.f.Reject an in-use indicator if the meter reading obtained differs from that of the reference standard bymore than two units in either direction.Figure 1-9b.Checking In-Use Field IndicatorsSECTION VII - DeletedChange 31-51