T.O. 33B-1-16-926.8.5 Density.Radiographic density is the blackening or darkening produced on the radiograph resulting from the metallic silverdeposits remaining on the film after exposure and processing. Density is measured in terms of visible lighttransmission. The accepted scale of density measurement is the logarithm of the ratio of incident light to transmittedlight as given by the following equation:D = log IIoEWhere:D = densityIo= intensity of incident lightIt = intensity of transmitted light6.8.5.1Measurement of radiographic density SHALL be done with electronic direct-reading type densitometers. The electronicdirect-reading type densitometer is more accurate than the visual type. This densitometer SHALL be capable ofmeasuring the light transmitted through a radiograph with a film density up to 4.0 with a density unit resolution of0.02. When film densities greater than 4.0 are required to perform a radiographic inspection a densitometer applicableto film densities up to maximum density is necessary. A photographic or radiographic-calibrated reference densitystrip, traceable to the National Institute of Standards and Technology (NIST), SHALL be used to calibrate thedensitometer prior to determining the density of a radiograph. These calibrated density strips shall be replacedwhenever they are physically damaged (i.e., scratched, crimped, or become wet by any fluid) to such an extent that itmay influence their effectiveness. The carbon, dot printed, etc. density strips SHALL NOT be used even though theymay be NIST traceable. These strips are not able to correlate the densitometer directly to Air Force radiographic needs.Each type of calibrated reference density strip will calibrate the densitometer to a different standard level. Therestrictive use of only the photographic or radiographic calibration reference density strip will better enable thestandardization of all densitometers to a single calibration value establishing a common (H and D units) density for agiven radiographic inspection. The aperture of the densitometer SHALL be black in color. If it is not, it may bedarkened with a black magic marker or other indelible ink.6.8.5.1.1While performing the densitometer calibration procedure, the following SHALL apply:a. Follow manufacturer's instructions, substituting the calibration strip supplied with the instrument withthe NIST traceable radiographic calibration reference density strip.b. The calibration reference density strip SHALL be removed from its protective cover during thecalibration procedure and maintained in its protective cover when not in use.c. The calibration reference density strip SHALL NOT be pulled or slid when it is between the apertureand stage diffuser. The aperture SHALL be raised so that it is not in contact with the density strip whenthe strip is being repositioned or removed from the densitometer.d. Calibrated reference density strip measurements SHALL be determined from the center of the steps thatare used for the calibration procedure.e. While reading the density of a radiograph, DO NOT pull or slide it between the aperture and stagediffuser. When repositioning or removing the radiograph from the densitometer the aperture SHALLbe raised.6.8.5.1.1.1The density of a radiograph is important. Densities less than 0.5 show very little of the object due to three factors: (1)the density of the emulsion base, (2) the basic "fog" of the film and (3) the lack of uniform response of the film at low
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business