T.O. 33B-1-16-55geometric unsharpness that can be tolerated sets the lower limit for the adjustable parameter. Further demands onsharpness are paid for in intensity. The unsharpness is inversely proportional to the source-object distance, whereas theintensity is inversely proportional to the square of this distance. Thus, the trade-off of intensity for sharpness is not anequitable one. Nonetheless, this uneven exchange is necessary in many cases because it is very important to achievegood geometric definition. The basic principles of shadow formation must be given primary consideration to insuresatisfactory sharpness and low distortion of radiographic images. Distortion cannot be entirely eliminated since someof the test object will be further away from the film than others and radiation from any source cannot be made ideallyparallel; images will always be less than perfect. In summary then, five general rules can be stated which promotequality assurance from geometric considerations.a. The X-rays should proceed from as small a local spot as other considerations will allow.b. The distance between the source and the object should be as great as practical.c. The film should be as close as possible to the object being radiographed.d. The central beam should be as nearly perpendicular to the film as possible.e. As far as the shape will allow, planes within the specimen plane of interest should be parallel with thefilm.6.7.2.6 FilmPlacement.After the film and film holder have been chosen, the film position in relation to the part must be considered. Inproduction radiography of small parts, this is a simple matter of laying the parts on the film holder. With complexstructures, film positioning is not usually as simple. A few rules can be of assistance in such inspection situations.a. Always position the film as close to the area of interest as possible.b. Attempt to locate the film so that the plane of the area of interest and the film are perpendicular to theradiation beam. This is to prevent distortion in the final image.6.7.2.6.1In positioning the film, care should be used to prevent sharp bends in the film or applying pressures to the film holderthat can produce pressure marks or crimp marks (artifacts) on the final image. In radiography of curve surfaces, thesource and film should be positioned, if possible, to take the best advantage of the inverse square law and to prevent asmuch distortion as possible. Flexible film holders should be used in order to place the film as near as possible to thesurface of the test object. It may be noted in Figure 6-25 that the distance from source to the entire surface of the film isnearly constant and the thickness of the test object is also a constant to the path of radiation. This preferred positioningis not always possible, but it should be used when practical.
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business