Table 3-2. Functional Test Procedures - Continued
Table 3-2. Functional Test Procedures - Continued

M140 Chemical Agent Alarm \(Nsn 6665-01-083-2749\)
Page Navigation
  41    42    43    44    45  46  47    48    49    50    51  
img
TM 3-6665-329-13&P
Table 3-2. Functional Test Procedures (Cont)
Test
No.
Test
Procedure
8.
Comparator Test
Adjust variable power supply to 20 volts dc.
e.
f.
Read multimeter. If it reads between 3.3
volts and 4.3 volts, continue testing. If
not, refer to Troubleshooting paragraph 3-7.4.
Rotate ALARM TEST switch to position 6.
g.
h.
Read multimeter. If it reads between 5.6
volts and 6.6 volts, continue testing. If
not, refer to Troubleshooting paragraph 3-7.4.
i.
Set multimeter to measure 50 volts dc scale.
Rotate ALARM TEST switch to position 10.
j.
k.
Read multimeter. If it reads between 9.4
volts and 10.4 volts, continue testing. If
not, refer to Troubleshooting paragraph 3-7.4.
1.
Rotate ALARM TEST switch to position 12.
m
Read multimeter. If it reads between 11.6
volts and 12.6 volts, continue testing. If
not, refer to Troubleshooting paragraph 3-7.4.
n.
Power OFF.
Remove variable power supply leads from sockets 7
o.
and 13 of extender board.
Remove multimeter leads.
p.
Remove extender board and place in accessory
q.
shelf.
3-25