TM 10-8415-208-12Table 3-2. Unit Preventive Maintenance Checks and Services for the STEPO-I(Cont)A - After OperationQ - QuarterlyINTERVALItemItem To BeProceduresEquipment IsNo.InspectedCheck For And Have RepairedNot Ready/AQOr Adjusted As NecessaryAvailable If:WARNINGCompressed oxygen hazard. Replacethe rebreather oxygen bottle if it issuspected to be leaking. Damagedoxygen bottles must be repaired ONLYby personnel trained and certified toaccomplish such work. Tampering withdamaged oxygen bottles could result inpersonnel injury or death due tocombustion or sudden release of highpressure.54Oxygen BottleEnsure bottle is pressurized andBottle is below 2,400indicates 2,400 PSIG and date of lastPSIG or date of lasthydrostatic test is current.hydrostatic test is notcurrent.NOTEThe hydrostatic test date is on thelabel or stamped on the cylindernear the neck.55Check hydrostatic test date.Test date is over 3years old.56AdjustableInspect webbing for rips, tears,Webbing is ripped orStrapsmissing or broken buckles andtorn or buckles arecleanliness.broken or missing.NOTEThe rebreather system shall be testedbefore initial wear, after each use,monthly, after specified maintenanceand when damage is suspected dueto a visual inspection or poorperformance.57BreathingPerform leak test. (SeeFails the leak test.System Leakparagraph 3-10.4.2.)Test58OxygenPerform oxygen constant flowFails the test.Constant Flowtest. (See paragraph 3-10.4.3)Rate TestIceVest(RebreatherSystem Mode)59BatteryEnsure a battery available.Battery is notavailable.Change 2 3-141